Page 19 of 28
RE-CALIBRATION – 5 Point EDIT
To edit the calibration points for a 5 Point re-calibration follow the
steps below.
NOTE: If the factory default values are acceptable, skip this
section and proceed to the re-calibration 5-Point START procedure.
Keystroke
Display
1. With unit OFF, press and
hold the MIN/MAX key,
turn the unit on using the
ON/OFF key, then release
Top line reads “RE-CAL”.
Bottom line reads “EDIT”.
2. Press the PRGM key.
Top line reads “CAL POINT
1”. Bottom line displays the
cal point value.
3. Press the or arrow
keys to edit the selected
digit. Use the or arrow
keys to change the cursor
position.
Note: For 0% go
directly to step 4.
Bottom line displays the cal
point value. The cursor flashes
at the first position while the
value is changed, then moves
to the right position when the
right arrow key is pressed.
4. Press the arrow key
while on the right most digit
to proceed.
Top line reads “CAL POINT
2”. Bottom line displays the
cal point value.
5. Repeat steps 3 and 4 for
CAL POINTS 2, 3, 4 and 5.
Top line reads “CAL POINT
2/3/4/5”. Bottom line displays
the cal point value.
6. After editing CAL
POINT 5 press the arrow
key while on the right most
digit to proceed.
Top line reads “RE-CAL”.
Bottom line reads “EDIT”.
7. To perform the 5-point
re-cal, press the arrow
key until START is
displayed on the bottom
line.
OR
To exit without performing
the 5-point re-cal press the
arrow key
Top line reads “RE-CAL”.
Bottom line, “START”.
Continue with 5-Point Re-
calibration procedure at step 3
on next page.
OR
Returns to Measure Mode.
RE-CALALIBRATION – 5 Point START
To begin the 5-point re-calibration procedure, turn the unit OFF
and follow the steps below.