Page 21 of 28
12. Press the arrow key.
Returns to Measure Mode.
RE-CALIBRATION – Restore Factory Defaults
To restore the re-calibration data to the factory defaults, follow
these steps:
Keystroke
Display
1. With unit OFF, press and
hold the MIN/MAX key,
turn the unit on using the
ON/OFF key, then release.
Top line reads “RE-CAL”.
Bottom line reads “EDIT”.
2. Press the arrow key
twice.
Top line reads “RE-CAL”.
Bottom line reads “RESTORE
DEFAULTS”.
3. Press the PRGM key.
Top line reads “RESTORE
DEFAULTS”.
Bottom reads “YES” or “NO”.
4. Use the and arrow
keys to select YES or NO
when asked to restore
defaults.
Top line reads “RESTORE
DEFAULTS”.
Bottom reads “YES” or “NO”.
5. Press the PRGM key at
YES to restore the Factory
Default Calibration data or
at NO to exit without
restoring.
Top line reads “RE-CAL”.
Bottom line reads “RESTORE
DEFAULTS”. Factory
defaults have been restored.
6. Press the arrow key.
Returns to Measure Mode.