background image

Page 17 of 28

Re-Calibration

The manometer can be re-calibrated in the field for zero, span, and 

linearity.  The proper primary standards must be available prior to 

calibrating  the  manometer.    These  standards  should  meet  the 

accuracy requirements for your company or industry. The factory 

follows the guidelines established by ANSI / NCSL Z540-1-1994 

which requires that the primary standard be 4 times more accurate 

than the unit under test.   

The  re-calibration  is  not  intended  to  replace  the  Factory  Lab 

Calibration Procedure.  It is intended to correct the curve fit if the 

actual sensor characteristics change slightly over time. 

For sensors up to 200 PSI, the factory recommends a ±0.0015% of 

reading  deadweight  tester.    For  sensors  200  PSI  and  above,  a 

±0.0030%  of  reading  deadweight  tester  is  recommended.    If 

calibrating  using  inches  of  water  units,  be  sure  to  match  the 

reference temperature of water in both the unit under test and the 

manometer. 

1-point (within upper 50% of Full Scale), 5-point (nominal values 

of 0%, 25%, 50%, 75% & 100% of Full Scale), and restore factory 

default  re-calibration  options  are  offered.  For  the  5-Point  re-

calibration,  points  2,  3  and  4  can  be  adjusted  within  ±1%  of 

reading around the nominal values. Point #5 can be adjusted within 

-1% of reading around nominal. Point #1 is fixed. 

For example:  for a 2000 inH2O sensor, Point # 2 (25%) can be 

edited  form  495  to  505  inH2O.    Point  #5  (100%)  can  be  edited 

from 1980 to 2000 inH2O. 

The  unit  can  only  be  re-calibrated  if  the  calibration  points  are 

within 5 times the accuracy of the original factory calibration (e.g. 

@ 0.05% accuracy, the point limit is ±0.25% of Full Scale).  If the 

re-calibration  procedure  generates  a  new  value  outside  this  limit 

the  procedure  will  fail.  In  this  case  the  unit  would  need  to  be 

returned to the factory for service. 

Once  a  re-calibration  has  been  performed  (either  1-point  or  5-

point)  the  unit  will  continue  to  allow  future  re-calibrations  only 

with  that  type  of  re-calibration.  In  order  to  enable  the  other  re-

calibration type, the user must first restore the re-calibration data to 

the factory defaults. 

Summary of Contents for HHP350 Series

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info User sGuide HHP350 SERIES High Accuracy Manometer Shop online at MADE IN...

Page 2: ...ged by the United Kingdom Office Toll Free 0800 466 342 TEL 33 0 161 37 29 00 FAX 33 0 130 57 54 27 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Toll Free 0 800...

Page 3: ...ressure sensors are supported see Specification section for supported types and pressure ranges Pressure can be displayed in selectable engineering units of measure Features include Min Max capture di...

Page 4: ...7 Flow Unit Select 8 Damp Rate Select 8 User Info Select 10 Auto Shut Off 11 Lockout Select 12 Header Name 13 Contrast Select 14 Data Logging 15 Leak Test 16 Re Calibration 17 RE CALIBRATION 1 Point E...

Page 5: ...Max function of the manometer When activated the minimum value is displayed on the upper left of the display and the maximum value on the upper right This key also deactivates and resets this functio...

Page 6: ...elects the programmable register to be edited with prompt for password if Lockout is set After the register has been edited pressing the PRGM key enters the new setting into the manometer s non volati...

Page 7: ...limit a ZERO RANGE ERROR message appears indicating that the procedure failed 3 Program Mode The program mode is used to configure the manometer for Measure Mode operation After the PRGM key is presse...

Page 8: ...PRGM key Top line reads PROGRAM MODE and bottom line reads UNITS SELECT 2 Press the PRGM key Top line reads UNITS SELECT and bottom line shows current engineering units 3 Press the or arrow key until...

Page 9: ...T 4 Press the PRGM key See note 1 at bottom of this table Top line reads VALUE Bottom line reads CHANGE YES 5 Press the PRGM key to change the value Top line reads USER UNIT VALUE 6 Start entering the...

Page 10: ...or L min The primary element must be a differential pressure square root type device such as a pitot tube orifice plate or venturi The flow constant and flow units description are programmed into the...

Page 11: ...ads PROGRAM MODE and bottom line reads UNITS SELECT 2 Press the arrow key Bottom line reads DAMP RATE SELECT 3 Press the PRGM key Top line reads DAMP RATE SELECT 4 Press the or arrow key until the des...

Page 12: ...Press the arrow key twice Bottom line changes to USER INFO SELECT 3 Press the PRGM key Bottom line shows serial number 4 Press the arrow key Software version number shown 5 Press the arrow key Manufa...

Page 13: ...n the User Info Select table Top line reads AUTO SHUT OFF and bottom line reads ENTER TO SELECT 2 Press the PRGM key then the or arrow keys until the desired shut off time is shown Top line reads AUTO...

Page 14: ...re Mode press the PRGM key If the Lockout is set enter the correct password when prompted Top line reads PROGRAM MODE and bottom line reads UNITS SELECT 2 Press the arrow key twice Bottom line reads U...

Page 15: ...ds PROGRAM MODE and bottom line reads USER INFO SELECT 6 Press the arrow key Returns to Measure Mode 7 Press the or arrow keys to set the correct alpha numeric value Displays a number between 0 and 9...

Page 16: ...ee times Bottom line reads CONTRAST SELECT 3 Press the PRGM key Top line reads CONTRAST SELECT and bottom line shows a numerical value 4 Press the or arrow keys to increase or decrease the contrast va...

Page 17: ...s RECORD 4 Press the PRGM key Top line reads RECORD MODE and bottom line reads AUTO or MANUAL 5 Press the PRGM key at AUTO to start automatic logging or at MANUAL to start manual logging mode Top line...

Page 18: ...ottom line reads LEAK TEST 3 Press the PRGM key Top line reads LEAK TEST and bottom line reads CONFIGURE 4 Press the PRGM key Top line reads Leak Test Period bottom X X MIN 5 Use the keys to input tes...

Page 19: ...ometer 1 point within upper 50 of Full Scale 5 point nominal values of 0 25 50 75 100 of Full Scale and restore factory default re calibration options are offered For the 5 Point re calibration points...

Page 20: ...PRGM key Top line reads CAL POINT and bottom line displays the cal point value 5 Press the or arrow keys to edit the selected digit Use the or arrow keys to change the cursor position Value entered m...

Page 21: ...ition while the value is changed then moves to the right position when the right arrow key is pressed 4 Press the arrow key while on the right most digit to proceed Top line reads CAL POINT 2 Bottom l...

Page 22: ...t takes new zero Top line reads POINT 1 ZERO Bottom line displays live applied pressure POINT 1 has been taken 7 Press the arrow key while on the right most digit to proceed Top line reads POINT 2 APP...

Page 23: ...Press the arrow key twice Top line reads RE CAL Bottom line reads RESTORE DEFAULTS 3 Press the PRGM key Top line reads RESTORE DEFAULTS Bottom reads YES or NO 4 Use the and arrow keys to select YES o...

Page 24: ...lity hysteresis and resolution Warm up time 5 minutes Zero unit at working ambient temperature before use Temperature Storage 40 C to 60 C 40 F to 140 F Operating 5 C to 50 C 23 F to 122 F Media Compa...

Page 25: ...itting Do not tighten the fitting without using a wrench on the pressure manifold Failure to use a wrench on the manifold will damage the plastic enclosure and void warranty No torque should be applie...

Page 26: ...astic enclosure while installing 1 8 NPT fittings Changing the Batteries Adherence to the Specifications and Certification Safety Warnings sections of this manual shall be enforced when changing batte...

Page 27: ...d offs molded into the side of the compartment When a battery is installed with the polarity reversed the stand offs prevent the negative battery terminal from contacting the positive terminal in the...

Page 28: ...locations identified as P1 and P2 One location is the top of the keypad shown on page 1 The second marking is stamped into the pressure connection fitting next to the pressure connections as shown he...

Page 29: ...this sort of damaged occurs the unit must be returned to the factory for sensor replacement WARNING User must use a wrench on the pressure manifold when installing user s 1 8 NPT fitting Do not tighte...

Page 30: ...st be returned to the factory for servicing In this case contact the factory representative in your area or call the factory at the numbers listed below for a Return Material Authorization RMA number...

Page 31: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 32: ...heel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipment DAT...

Reviews: