background image

6

MAINTENANCE

Verify Accuracy

Before use, check the turbine’s accuracy and 

verify calibration.
1. Make sure there is no air in the system.
2. Measure an exact known volume into an

accurate container.

3. Verify the volume against the readout or

recording equipment.

For best results, accuracy should be verified 

periodically as part of a routine maintenance 

schedule.

Remove the Turbine

1. Ensure all liquid is drained from the tur-

  bine. Wear protective clothing as necessary.
2. Loosen both ends of the turbine. Use a

wrench only on the turbine’s wrench flats.

3. If the turbine is not immediately installed

again, cap lines as necessary.

Clean the Turbine

During use, the turbine should be kept full 

of liquid to ensure that drying does not occur 

inside the turbine. If drying or caking should 

occur, the rotor will stick or drag, affecting 

accuracy. To determine if the rotor is stuck or 

dragging, gently blow air through the meter 

and listen for the quiet whir of the rotor.

During turbine removal, liquid may spill. 

Follow the liquid manufacturer’s safety 

precautions for clean up of minor spills.

WARNING

CAUTION

Never blow compressed air through the 

meter. It could damage the rotor.

Follow the liquid manufacturer’s in-

structions for the disposal of contami-

nated cleaning solvents.

WARNING

1. Remove the turbine from the system fol-

lowing the directions above.

2. When the rotor turns freely, assemble and

install it again following the instructions

above.

Display Electronics

The  display  electronics  are  powered  by 

lithium batteries which provide at least four 

years of actual use. If the meter’s readout 

should become dim, blank or the low battery 

message appears (see below), the batteries 

should be replaced. Replacement batteries 

can be ordered from the factory. See details 

in the Parts Section.

When batteries are disconnected or fail, 

values in Batch and Cumulative Totals will 

remain. Factory and Field Calibration Curves 

are retained in the meter’s computer when 

power is lost.
It is strongly recommended that battery check 

and terminal cleaning be a part of a routine 

maintenance schedule. Battery terminals 

should be cleaned annually. Batteries can be 

replaced without removing the meter from the 

piping system.

Summary of Contents for FTB380 Series

Page 1: ...FTB380 SERIES Aluminum Turbine Flowmeters...

Page 2: ......

Page 3: ...Turbine Meter System prior to installation and use If you need assistance contact the Omega Flow Application Department This symbol is used throughout the manual to call your attention to safety messa...

Page 4: ...ine housing causing an internal rotor to spin As the rotor spins an electrical signal is generated in the pickup coil The electrical signal provides the output necessary to operate the on board displa...

Page 5: ...28 Standard Sieve 500 m Alternative Sieve No 35 All Series FTB380 turbines are Factory Mu tualApproved and carry a Class 1 Division 1 Approval for hazardous environments They are tested and calibrate...

Page 6: ...ext rollover occurs the computer will highlight an X100 icon This indicates to the operator that two zeroes must be added to the 6 digits shown Press the DISPLAY button briefly to switch between the T...

Page 7: ...a non standard application Field Calibration Procedures Dispense Display Method 1 Tofieldcalibrate pressandholdtheCALI BRATE and DISPLAY buttons for about 3secondsuntilyouseeFLdCAL Release bothbuttons...

Page 8: ...l Follow the liquid manufacturer s safety precautions for clean up of minor spills WARNING CAUTION Never blow compressed air through the meter It could damage the rotor Follow the liquid manufacturer...

Page 9: ...ery terminals and remove any corrosion 4 Install the new batteries and make sure the positive posts are positioned correctly When the batteries are installed correctly the computer powers on automatic...

Page 10: ...s to Check all electrical connections Reference recording device appropriate installation instructions 10 Accuracy needs verification Complete normal accuracy verification procedures Repeat periodical...

Page 11: ...1 in 2 in Weight with Computer Electronics Pounds 1 35 lbs 1 35 lbs 3 0 lbs Kilograms 61 kg 61 kg 1 36 kg Performance Linear Range for 1 in Low Flow N A Linear Range for 1 in 1 5 of reading Linear Ra...

Page 12: ...the computer electronics Dimensions A C B Display Specifications Input Pulse Rate Minimum Pulse In DC Minimum Coil Input 10 Hz Maximum Raw 1 000 Hz K Factor Minimum 01 pulses unit Maximum 999 999 pul...

Page 13: ...ctor output Except as noted Options and Accessories are for Display Models only and only one Module may be installed per unit All models are supplied with 10 ft cable Cable may be cut or extended as r...

Page 14: ......

Page 15: ......

Page 16: ...M 5065 1111...

Reviews: