Omega FTB380 Series User Manual Download Page 6

If for any reason the display electronics 

need to be mounted on your turbine, simply 

mount the display on the turbine with the four 

screws at the corners of the faceplate. Make 

sure the seal is fully seated before tightening 

the screws.
If you ordered the FTB380 Series turbine with 

an accessory module, please review and thor-

oughly understand all installation instructions 

before proceeding.
Avoid electronically “noisy” environments. 

Install at least 6 inches (15.2 cm) away from 

motors, relays, or transformers.

4

OPERATION

All operations are reflected in the LCD read-

out. The large center digits indicate amounts, 

where smaller words or “icons” located above 

and  below  indicate  specific  information 

regarding totals, flow, calibration and units 

of measure.

Activate the Meter

The computer is on continuously and always 

ready to perform. The computer is powered 

by field replaceable batteries. When display 

becomes dim, faded or the low battery mes-

sage appears (see below), the batteries need 

to be replaced. Reference the Maintenance 

Section for details.

Batch and Cumulative Totals

The computer maintains two totals.The Cumu-

lative Total provides continuous measurement 

and cannot be manually reset. The Batch Total 

can be reset to measure flow during a single 

use. The Cumulative Total is labeled TOTAL 

1, Batch Total is labeled TOTAL 2 BATCH.
When the Cumulative Total reaches a display 

reading of 999,999 the computer will highlight 

an X10 icon. This indicates to the operator that 

a zero must be added to the 6 digits shown. 

When the next rollover occurs, the computer 

will highlight an X100 icon. This indicates to 

the operator that two zeroes must be added to 

the 6 digits shown.

Press the DISPLAY button briefly to switch 

between  the TOTAL  1, TOTAL  2  BATCH 

and  FLOWRATE.  Press  DISPLAY  briefly 

to display the TOTAL 2 BATCH. Hold the 

DISPLAY button for 3 seconds to reset the 

Batch Total to zero.
When fluid is flowing through the meter, a 

small propeller icon is highlighted.

Flowrate Feature

To use this feature, press and release DIS-

PLAY button until FLOWRATE icon appears. 

The factory set time base will be highlighted 

to the right of FLOWRATE (M = Minutes, 

H = hours, D = days). When FLOWRATE is 

invoked, the display will be indicating rate 

of flow.

Factory and Field Calibration

All calibration information is visible to the 

user as icons on the top line of the display, 

above the numeric digits.
All  units  are  configured  with  a  “factory” 

calibration. Both gallons and litres are avail-

able (“GL” or “LT” will be displayed). While 

holding the CALIBRATE button, briefly press 

DISPLAY to toggle between gallons and litres. 

This factory calibration (indicated with FAC) 

is permanently programmed into the computer 

and is not user adjustable.
NOTE:  Your computer may have other units of 

measure programmed into it. If so, holding 

the CALIBRATE button and momentarily 

pressing the DISPLAY button will toggle 

through all factory set units. Other pos-

sible units are: IGL (imperial gallon), QT 

(quart), CF (cubic feet), CM (cubic meter), 

BL (42 gal. barrel), CC (cubic centimeter) 

or OZ (ounce).

Switching  between  different  units  will  not 

corrupt the Total’s contents. For example, in 

GL mode, the computer totalizes 10.00 gallons, 

if the user switches to LT mode, the display 

will read 37.85 litres (the same volume, dif-

ferent unit).

Summary of Contents for FTB380 Series

Page 1: ...FTB380 SERIES Aluminum Turbine Flowmeters...

Page 2: ......

Page 3: ...Turbine Meter System prior to installation and use If you need assistance contact the Omega Flow Application Department This symbol is used throughout the manual to call your attention to safety messa...

Page 4: ...ine housing causing an internal rotor to spin As the rotor spins an electrical signal is generated in the pickup coil The electrical signal provides the output necessary to operate the on board displa...

Page 5: ...28 Standard Sieve 500 m Alternative Sieve No 35 All Series FTB380 turbines are Factory Mu tualApproved and carry a Class 1 Division 1 Approval for hazardous environments They are tested and calibrate...

Page 6: ...ext rollover occurs the computer will highlight an X100 icon This indicates to the operator that two zeroes must be added to the 6 digits shown Press the DISPLAY button briefly to switch between the T...

Page 7: ...a non standard application Field Calibration Procedures Dispense Display Method 1 Tofieldcalibrate pressandholdtheCALI BRATE and DISPLAY buttons for about 3secondsuntilyouseeFLdCAL Release bothbuttons...

Page 8: ...l Follow the liquid manufacturer s safety precautions for clean up of minor spills WARNING CAUTION Never blow compressed air through the meter It could damage the rotor Follow the liquid manufacturer...

Page 9: ...ery terminals and remove any corrosion 4 Install the new batteries and make sure the positive posts are positioned correctly When the batteries are installed correctly the computer powers on automatic...

Page 10: ...s to Check all electrical connections Reference recording device appropriate installation instructions 10 Accuracy needs verification Complete normal accuracy verification procedures Repeat periodical...

Page 11: ...1 in 2 in Weight with Computer Electronics Pounds 1 35 lbs 1 35 lbs 3 0 lbs Kilograms 61 kg 61 kg 1 36 kg Performance Linear Range for 1 in Low Flow N A Linear Range for 1 in 1 5 of reading Linear Ra...

Page 12: ...the computer electronics Dimensions A C B Display Specifications Input Pulse Rate Minimum Pulse In DC Minimum Coil Input 10 Hz Maximum Raw 1 000 Hz K Factor Minimum 01 pulses unit Maximum 999 999 pul...

Page 13: ...ctor output Except as noted Options and Accessories are for Display Models only and only one Module may be installed per unit All models are supplied with 10 ft cable Cable may be cut or extended as r...

Page 14: ......

Page 15: ......

Page 16: ...M 5065 1111...

Reviews: