background image

5

CALIBRATION

Verify Accuracy Before 

Beginning Field Calibration

For the most accurate results, dispense at a 

flowrate which best simulates your actual oper-

ating conditions. Avoid “dribbling” more fluid 

or repeatedly starting and stopping the flow. 

This can result in less accurate calibrations.
Make sure you meet the meter’s minimum 

flowrate requirements:

A109GMA025NA1

 mete

r

.3 GPM (1.1 LPM) 

A109GMA100NA1

 meter

 

3 GPM (11.0 LPM)

 

A109GMA200NA2

 meter 

 

30 GPM (113 LPM)

The use of a uniformly dependable, accurate 

calibration container is recommended for the 

most accurate results. Due to high flowrate, 

it is strongly recommended that calibration 

be completed with a combination of volume 

and weight using fine resolution scales. For 

best results, the meter should be installed and 

purged of air before field calibration.

Field Calibration with 

Computer Display

Field  Calibration  and  Factory  Calibration 

are defined in the Operation Section. Factory 

calibration settings are programmed into each 

computer during manufacturing, using stod-

dard test solvent at 70° F (21° C). Settings 

are  correct  for  light  liquids  such  as  water, 

gasoline or diesel. Readings using the Factory 

Calibration (FAC) may not be accurate in some 

situations, for example, “heavy” liquids such 

as motor oil under extreme temperature condi-

tions, non-standard plumbing configurations 

or with fluids other than water.
For improved accuracy under such conditions, 

the computer allows for “field” calibration, that 

is, user entry of custom calibration parameters. 

A “single point” calibration may yield accept-

able accuracy when used in a non-standard 

application.

Field Calibration Procedures

(Dispense/Display Method)

1. To field calibrate, press and hold the CALI-

BRATE and DISPLAY buttons for about

3 seconds until you see FLdCAL. Release

both buttons and you will see dd000.0. You

are now in the field calibration mode.

2. Dispense a known amount of fluid at a

flowrate representative of the application.

Any  amount  between  000.1  and  999.9

units can be used. Display will count up

while fluid is flowing through the meter.

3. The DISPLAY button can then be pushed

to select the digit location and the CALI-

BRATE button can be pushed to scroll

the desired value at the blinking position.

Edit the amount shown with the value that

was dispensed above. Values from 000.1

to 999.9 can be entered.

4. When satisfied with the value, press both

CALIBRATE and DISPLAY buttons si-

multaneously. CALEnd will be displayed

and unit will go back to normal operation,

less the FAC (factory calibration) icon.

5. The meter will now be operating with a

custom calibration number unique to the

above  dispense  procedure.  No  unit  of

measure (gallon, litre, etc.) icon will be

highlighted

NOTE:  To return to factory calibration (FAC), 

press and hold both CALIBRATION and 

DISPLAY buttons for about 3 seconds, 

until FAcCAL is displayed. Then release 

buttons.  Unit  should  return  to  normal 

operation and FAC icon visible

NOTE:  If the field calibration mode is entered 

and NO fluid is dispensed, then upon leav-

ing, the computer will use data from the 

last successful field calibration.

The “field” calibration may be set by the user, 

and can be changed or modified at any time 

using the calibration procedure described in 

the  Calibration  Section.  Totals  or  flowrate 

derived from the field calibration are invoked 

when the FAC icon is no longer visible on the 

top line of the display.

Summary of Contents for FTB380 Series

Page 1: ...FTB380 SERIES Aluminum Turbine Flowmeters...

Page 2: ......

Page 3: ...Turbine Meter System prior to installation and use If you need assistance contact the Omega Flow Application Department This symbol is used throughout the manual to call your attention to safety messa...

Page 4: ...ine housing causing an internal rotor to spin As the rotor spins an electrical signal is generated in the pickup coil The electrical signal provides the output necessary to operate the on board displa...

Page 5: ...28 Standard Sieve 500 m Alternative Sieve No 35 All Series FTB380 turbines are Factory Mu tualApproved and carry a Class 1 Division 1 Approval for hazardous environments They are tested and calibrate...

Page 6: ...ext rollover occurs the computer will highlight an X100 icon This indicates to the operator that two zeroes must be added to the 6 digits shown Press the DISPLAY button briefly to switch between the T...

Page 7: ...a non standard application Field Calibration Procedures Dispense Display Method 1 Tofieldcalibrate pressandholdtheCALI BRATE and DISPLAY buttons for about 3secondsuntilyouseeFLdCAL Release bothbuttons...

Page 8: ...l Follow the liquid manufacturer s safety precautions for clean up of minor spills WARNING CAUTION Never blow compressed air through the meter It could damage the rotor Follow the liquid manufacturer...

Page 9: ...ery terminals and remove any corrosion 4 Install the new batteries and make sure the positive posts are positioned correctly When the batteries are installed correctly the computer powers on automatic...

Page 10: ...s to Check all electrical connections Reference recording device appropriate installation instructions 10 Accuracy needs verification Complete normal accuracy verification procedures Repeat periodical...

Page 11: ...1 in 2 in Weight with Computer Electronics Pounds 1 35 lbs 1 35 lbs 3 0 lbs Kilograms 61 kg 61 kg 1 36 kg Performance Linear Range for 1 in Low Flow N A Linear Range for 1 in 1 5 of reading Linear Ra...

Page 12: ...the computer electronics Dimensions A C B Display Specifications Input Pulse Rate Minimum Pulse In DC Minimum Coil Input 10 Hz Maximum Raw 1 000 Hz K Factor Minimum 01 pulses unit Maximum 999 999 pul...

Page 13: ...ctor output Except as noted Options and Accessories are for Display Models only and only one Module may be installed per unit All models are supplied with 10 ft cable Cable may be cut or extended as r...

Page 14: ......

Page 15: ......

Page 16: ...M 5065 1111...

Reviews: