Symptom
Possible Cause
Method of Correction
Flow indication is not
linear and output too low
at lower flows
Attempting to measure
flows below the specified
minimum for the unit
Liquid is too viscous
Entrapped air or gas
Mechanical sensor
assembly damaged
Use higher flow rates
Best results are obtained
with liquids that have a
viscosity of 10cS or less.
Contact the Omega
Engineering Service
Department for further
information
Remove entrapped air
(see section C2)
Unit must be returned for
repair (see Section D3)
Flow reading accuracy is
outside the specification
Liquid has a different
viscosity and / or specific
gravity than water at 20°C
Entrapped air or gas
Mechanical sensor
assembly damaged
Calculate the flow reading
error / offset as explained
in section C6
Remove the entrapped air
(see section C2)
Unit must be returned for
repair (see Section D3)
Flow indication is fixed at
one value
The maximum rated flow
for the unit is being
exceeded.
Output circuits may be
damaged
Reduce the flow
Unit must be returned for
repair (see Section D3)
Summary of Contents for FLR 1000
Page 1: ...FLR 1000 1000BR 1000ST Series Flow Sensors Meters For Liquids...
Page 2: ......
Page 25: ...pg 25 of 33 FLR 1000 BR FLR 1000 ST For Liquids 1 4 Compression Fittings Shown...
Page 26: ...pg 26 of 33 FLR 1000 ST I For Liquids 1 4 Fittings Shown...
Page 27: ...pg 27 of 33 FLR 1000 D For Liquids 1 4 Acetal Fittings Shown...
Page 28: ...pg 28 of 33 FLR 1000 BR D FLR 1000 ST D For Liquids 1 4 Compression Fittings Shown...
Page 32: ...pg 32 of 33...
Page 33: ...M4574 0807 pg 33 of 33...