The FLR 1000-D, 1000-ST-D and 1000-BR-D series feature an integral
display that provides a local flow reading. These units also have a 0-5VDC
analog output available. If required, this may be connected to another
display, data acquisition system or voltmeter.
A stable D.C. power supply is required to operate the unit. The voltage
and current requirements depend on the configuration of the unit. Full
details may be found in the Specification section of this manual.
Connecting wires should be as short as possible to avoid voltage drops.
Twisted 2 pair conductor cable should be used if the length of the power
wires is to be longer than 1 meter.
Units are supplied with an integral 4 pin connector. Connections to the unit
are made using a mating cable assembly or power adapter package as
detailed in the following sections (parts 5.b to 5.f of this manual). A
connector pin and wire color cross reference may also be found in
Appendix G (Page 29) of this manual.
b)
Connecting the Cable Assembly
The connector on the end of the cable assembly should be pushed into the
mating socket on the sensor taking care to ensure that it is the correct
way up.
Connecting the Cable Assembly
(FLR 1000 shown, other models similar)
Electrical Connections are made to the cable assembly as detailed in the
following sections.
Summary of Contents for FLR 1000
Page 1: ...FLR 1000 1000BR 1000ST Series Flow Sensors Meters For Liquids...
Page 2: ......
Page 25: ...pg 25 of 33 FLR 1000 BR FLR 1000 ST For Liquids 1 4 Compression Fittings Shown...
Page 26: ...pg 26 of 33 FLR 1000 ST I For Liquids 1 4 Fittings Shown...
Page 27: ...pg 27 of 33 FLR 1000 D For Liquids 1 4 Acetal Fittings Shown...
Page 28: ...pg 28 of 33 FLR 1000 BR D FLR 1000 ST D For Liquids 1 4 Compression Fittings Shown...
Page 32: ...pg 32 of 33...
Page 33: ...M4574 0807 pg 33 of 33...