Model DFG-RS5 Digital Force / Torque Indicator
User’s Guide
26
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
Causes:
1. The calibration weight does not match the set value.
2. If using a DFG-RSA configurable adapter, ensure that the output signal leads (SG+ and SG-)
have been installed into the appropriate terminal blocks. Some sensor manufacturers
consider SG+ to be a compression value, while others consider it to be a tension value. If the
indicator is expecting a compression load but it is receiving a tension signal, calibration
cannot continue. Verify that the tension/compression indicator on the home screen properly
corresponds to the load direction, and switch the signal leads, if required.
The entered calibration point is too close to the previous point.
11 PASSWORDS
Two separate passwords may be set to control access to the Calibration section and to the menu and
other keys. To access the passwords setup screen, select
Passwords
from the menu. The display
appears as follows:
Select “+ More” for additional options:
CALIBRATION
TENSION
Load too close
to previous.
Please try again.
CALIBRATION
Load not stable.
Please try again.
CALIBRATION
COMPRESSION
Load too low.
Please try again.
PASSWORDS
Calibration
Menu Key
Sensor
Lock
Units Key
Mode Key
+ More