10   CALIBRATION  

 

10.1 Initial Physical Setup 

The sensor should be mounted vertically to a test stand or fixture rugged enough to withstand a load 
equal to the full capacity of the sensor. Certified deadweights, torque arms/wheels, and/or master load 
cells should be used, along with appropriate mounting brackets and fixtures. Caution should be taken 
while handling such equipment.  

 
10.2 Calibration Procedure 

In the interests of simplicity and brevity, the following instructions use force terminology only. Such 
wording is displayed only when a force sensor is being calibrated. When a torque sensor is being 
calibrated, the terms 

COMPRESSION

 and 

TENSION

 are replaced by 

CLOCKWISE

 and 

COUNTER-

CLOCKWISE

, respectively.  

 

1. Select 

Calibration

 from the menu. The display appears as follows: 

 

 
The sensor can be calibrated at up to 10 points in each direction. Enter the number of calibration 
points for each direction (compression and tension or clockwise and counter-clockwise). At least 
one point must be selected for each direction. For single-direction sensors such as Omega’s 
Series RLC02, only one direction is allowed.  
 

Note:

 To achieve the accuracy specification of ±0.2% + sensor, it is recommended to calibrate 

the sensor at 5 or more even increments in both the tension and compression directions. For 
example, a sensor with capacity of 10 lbF should be calibrated at 2, 4, 6, 8, and 10 lbF loads in 
each direction. 

 

2.  To escape the 

Calibration

 menu at any time, press 

ESCAPE

. The display appears as follows: 

 

 

Selecting “Cancel” will revert back to the Calibration setup. Selecting “Exit w/o saving” will return 
to the menu without saving changes. 

 

3.  After the number of calibration points has been entered, press 

ENTER

. The display appears as 

follows: 

CALIBRATION 

NOT COMPLETE 

 
   

Cancel

 

   Exit w/o saving 

CALIBRATION 

Enter # cal points  

(1 to 10) 

Compression: 

Tension: 

Summary of Contents for DFG-RS3

Page 1: ...Model DFG RS3 Digital Force Torque Indicator...

Page 2: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 1...

Page 3: ...provides setup safety and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions please do not hesitate to contact us Our techn...

Page 4: ...indicator and connected to another without the need for re calibration or re configuration All such data is saved within a PCB located inside the smart connector The model number serial number and ca...

Page 5: ...connected to a DFG RS3 indicator Resolution information is shown in the sensors user s guide 1 4 Safety Proper Usage Read through the following safety instructions thoroughly before using the DFG RS3...

Page 6: ...f suitably rugged materials and components Similar precautions should be taken when mounting the sensor to a test stand work bench or other piece of equipment 2 POWER The DFG RS3 is powered either by...

Page 7: ...into place with a click Fig 3 1 Appropriate orientation of the connector To release the connector press both buttons on either side of the indicator housing to release the sensor as shown in Fig 3 2 b...

Page 8: ...Primary reading The current displayed load reading See Operating Modes section for details If a sensor is not plugged in this value will be replaced by a message as follows SENSOR NOT CONNECTED 4 Load...

Page 9: ...upper load limit the displayed value is between the load limits the displayed value is less than the lower load limit 9 Set points The programmed load limit values Typically used for pass fail type te...

Page 10: ...e As described above the Plug TestTM smart connector retains all configuration and calibration data for the sensor which includes menu settings As such a sensor must be connected in order for menu cha...

Page 11: ...corner of the display will be blank When set points are enabled the following indicators are shown to the left of the primary reading the displayed value is greater than the upper load limit NO GO HIG...

Page 12: ...selected 9 COMMUNICATIONS AND OUTPUTS Communication with the DFG RS3 is achieved through the micro USB port located along the left side of the housing as shown in the illustration in the Power section...

Page 13: ...ssion clockwise values have positive polarity tension counter clockwise values have negative polarity Numeric Only Output format includes the value only Polarity same as above Invert Polarity Compress...

Page 14: ...nsor can be calibrated at up to 10 points in each direction Enter the number of calibration points for each direction compression and tension or clockwise and counter clockwise At least one point must...

Page 15: ...quired Do not yet attach any weights or apply any calibration loads Press ENTER 6 The display appears as follows Optionally exercise the sensor several times at full scale if possible then press ENTER...

Page 16: ...F capacity sensor is calibrated and 5 data points were selected the load values will default to 10 20 30 40 and 50 lb Apply the calibration load Then press ENTER Repeat the above step for the number o...

Page 17: ...pter ensure that the output signal leads SG and SG have been installed into the appropriate terminal blocks Some sensor manufacturers consider SG to be a compression value while others consider it to...

Page 18: ...sed 11 2 Backlight There are several available initial settings applicable upon powering on the indicator To access these settings select Backlight from the menu The display appears as follows Selecti...

Page 19: ...functions for which audible tones will apply select Tones from the menu The screen appears as follows 11 5 Initial Mode This section is used to configure the initial mode upon powering on the indicat...

Page 20: ...nformation Welcome Screen The following screen is displayed at power up and can be accessed at any time by selecting Information from the menu RESTORE DEFAULT SETTINGS No Yes Digital Indicator Model D...

Page 21: ...0 of full scale display shows OVER at 110 and above Weight 0 7 lb 0 3 kg Included accessories Carrying case AC adapter battery USB cable resource CD USB driver and user s guide certificate of conforma...

Page 22: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 21 12 3 Dimensions IN MM...

Page 23: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 22...

Page 24: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 23 M 5250 1017...

Reviews: