background image

Model DFG-RS3 Digital Force/Torque Indicator 

 

User’s Guide 

 
 

 

 

 

10

 

6   SET POINT INDICATORS 

 

6.1 General Information 

Set points are useful for tolerance checking (pass/fail). Two limits, high and low, are specified and stored 
in the non-volatile memory of the instrument and the primary reading is compared to these limits.  
 

6.2 Configuration 

To configure set points, select 

Set Points

 from the menu. The screen appears as follows: 

 

 

 
Either one, two, or none of the set points may be enabled. To toggle between the tension and 
compression (or clockwise and counter-clockwise) directions, press the 

DIRECTION

 key. 

 
If two set points have been enabled, they are displayed in the upper left corner of the display. If only one 
set point has been enabled, the word “OFF” appears in place of the value. If no set points have been 
enabled, the upper left corner of the display will be blank.  
 
When set points are enabled, the following indicators are shown to the left of the primary reading: 

 
 

  – the displayed value is greater than the upper 

load limit (NO GO HIGH) 

 

  – the displayed value is between the limits (GO) 

 

 – the displayed value is less than the lower load 

limit (NO GO LOW) 

 
 

 

Note:

 Set point indicators reference the displayed reading, not necessarily the current live load. 

 
 

7   OPERATING MODES 

 

Caution! 

In any operating mode, if the capacity of the instrument has been exceeded by more than 110%, 
the display will show “OVER” to indicate an overload. A continuous audible tone will be sounded 
(if beeps are enabled) until the MENU key has been pressed or the load has been reduced to a 
safe level. 

 
Three operating modes are possible with the DFG-RS3 indicator. To cycle between the modes, press 

MODE

 while in the home screen. 

 
7.1 Real time (RT) 

The primary reading corresponds to the live measured reading.  
 

SET POINTS 

  Upper Disabled 
* Upper Enabled 

           

 

5.00 

lbFin 

   Lower Disabled 
* Lower Enabled 

            3.50 lbFin 

Summary of Contents for DFG-RS3

Page 1: ...Model DFG RS3 Digital Force Torque Indicator...

Page 2: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 1...

Page 3: ...provides setup safety and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions please do not hesitate to contact us Our techn...

Page 4: ...indicator and connected to another without the need for re calibration or re configuration All such data is saved within a PCB located inside the smart connector The model number serial number and ca...

Page 5: ...connected to a DFG RS3 indicator Resolution information is shown in the sensors user s guide 1 4 Safety Proper Usage Read through the following safety instructions thoroughly before using the DFG RS3...

Page 6: ...f suitably rugged materials and components Similar precautions should be taken when mounting the sensor to a test stand work bench or other piece of equipment 2 POWER The DFG RS3 is powered either by...

Page 7: ...into place with a click Fig 3 1 Appropriate orientation of the connector To release the connector press both buttons on either side of the indicator housing to release the sensor as shown in Fig 3 2 b...

Page 8: ...Primary reading The current displayed load reading See Operating Modes section for details If a sensor is not plugged in this value will be replaced by a message as follows SENSOR NOT CONNECTED 4 Load...

Page 9: ...upper load limit the displayed value is between the load limits the displayed value is less than the lower load limit 9 Set points The programmed load limit values Typically used for pass fail type te...

Page 10: ...e As described above the Plug TestTM smart connector retains all configuration and calibration data for the sensor which includes menu settings As such a sensor must be connected in order for menu cha...

Page 11: ...corner of the display will be blank When set points are enabled the following indicators are shown to the left of the primary reading the displayed value is greater than the upper load limit NO GO HIG...

Page 12: ...selected 9 COMMUNICATIONS AND OUTPUTS Communication with the DFG RS3 is achieved through the micro USB port located along the left side of the housing as shown in the illustration in the Power section...

Page 13: ...ssion clockwise values have positive polarity tension counter clockwise values have negative polarity Numeric Only Output format includes the value only Polarity same as above Invert Polarity Compress...

Page 14: ...nsor can be calibrated at up to 10 points in each direction Enter the number of calibration points for each direction compression and tension or clockwise and counter clockwise At least one point must...

Page 15: ...quired Do not yet attach any weights or apply any calibration loads Press ENTER 6 The display appears as follows Optionally exercise the sensor several times at full scale if possible then press ENTER...

Page 16: ...F capacity sensor is calibrated and 5 data points were selected the load values will default to 10 20 30 40 and 50 lb Apply the calibration load Then press ENTER Repeat the above step for the number o...

Page 17: ...pter ensure that the output signal leads SG and SG have been installed into the appropriate terminal blocks Some sensor manufacturers consider SG to be a compression value while others consider it to...

Page 18: ...sed 11 2 Backlight There are several available initial settings applicable upon powering on the indicator To access these settings select Backlight from the menu The display appears as follows Selecti...

Page 19: ...functions for which audible tones will apply select Tones from the menu The screen appears as follows 11 5 Initial Mode This section is used to configure the initial mode upon powering on the indicat...

Page 20: ...nformation Welcome Screen The following screen is displayed at power up and can be accessed at any time by selecting Information from the menu RESTORE DEFAULT SETTINGS No Yes Digital Indicator Model D...

Page 21: ...0 of full scale display shows OVER at 110 and above Weight 0 7 lb 0 3 kg Included accessories Carrying case AC adapter battery USB cable resource CD USB driver and user s guide certificate of conforma...

Page 22: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 21 12 3 Dimensions IN MM...

Page 23: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 22...

Page 24: ...Model DFG RS3 Digital Force Torque Indicator User s Guide 23 M 5250 1017...

Reviews: