6. Details of [Display] Menu
PV6253-F4E016
22
[Display wavelength][Full scale]
If
[Full scale]
is currently checked, the maximum indication range from 380 to 780 [nm]
will be displayed in a graph, in disregard of the set values of
[Min]
and
[Max]
. (Range from
440 to 780 [nm] on a model with 440nm - 840nm).
[Display wavelength][Short Wavelength]
Specify the short wavelength side of the indication range of measured data.
[Display wavelength][Long Wavelength]
Specify the long wavelength side of the indication range of measured data.
[Scale][Pitch]
Specify the spacing of graduations on the wavelength axis. The input range for this
specification is from 5 to 1000 [nm], and the graduations will be shown at the spacing value
entered.
6.8.3. [Reflectance] tab
The indication reflectance scale of measured data is set here.
[Display reflectance][Automatic scale]
If measurement is conducted with
[Automatic scale]
checked, the maximum reflectance
will be set in such a manner that the whole measured data will be displayed in a graph. The
unit is [%].
[Display reflectance][Max]
Specify the maximum value for the reflectance scale. The range of this setting is from 0.01
to 100 [%].
Summary of Contents for USPM-RUIII Series
Page 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Page 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Page 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Page 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Page 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...