3. Operating the centering telescope
PV6670-F4E003
7
3.
Operating the centering
telescope
For details on how to take measurements, see the Hardware and Software Operation Manuals for
the USPM-RUIII Lens Reflect meter.
(1)
First, start the system, conduct background and reference measurements, and then place the
lens to be inspected on the stage.
(2)
Bringing the measurement point to the surface of the lens under inspection
Look into the eyepiece (MICRO SWH10x) on the right side of the main unit and focus on the
surface of the lens being inspected, as shown in Figure 3.
Objective lens
Lens under
inspection
Face under inspection (the surface of
the lens under inspection)
Optical axis of the lens
under inspection
Optical axis of this
device
Apex
Measurement
light flux
Measurement
point
Figure 3. Measurement layout
Summary of Contents for USPM-RUIII Series
Page 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Page 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Page 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Page 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Page 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...