9. Details of [System] Menu
PV6253-F4E016
45
9.1.5. [File] tab
Specify a file data as the theoretical spectral reflectance of the reference. The file data will be a
reflectance in text format in steps of one [nm]. This allows that spectral reflectance data measured
with another spectrometer, etc., be used as the theoretical spectral reflectance data of the reference.
9.1.6. [Save] button
Saves the theoretical reflectance of the reference currently specified in the window to a file. For
instance, if a fixed value setting is currently specified and its value is 4.0 [%], the theoretical
spectral reflectance data of 4.0 [%] at all the measured wavelengths will be saved in the specified
file name.
Note: For details on the file format, see "12. File Configuration" described further ahead.
9.2. [Sampling] menu
Opens the
[Sampling setting]
window, where the basic parameters of the USPM spectrometer are
specified. Make setting for the quantity of light reading of the line sensor employed in the USPM
spectrometer. Quantity of light I read by the line sensor is represented by the following equation:
time)
Sampling
light x
of
(Quantity
=
∑
samplings
of
Number
I
The quantity of light of the line sensor is subjected to 12-bit A/D conversion. Therefore, if the
quantity of light sampled in one single sampling exceeds 12 bits (actually 4080), saturation will
occur, making it impossible to measure accurate values.
Note: If parameters have been set here, be sure to measure the background and reference once again.
Summary of Contents for USPM-RUIII Series
Page 42: ...PV6253 F4E016 USPM RU Ver3 21 004 Software Operation Manual...
Page 89: ...10 Help Menu PV6253 F4E016 47 10 Help Menu 10 1 Version menu Displays the version...
Page 92: ...11 About the Calibration Method for Wavelength Direction PV6253 F4E016 50...
Page 113: ...PV9386 F4E005 1 USPM RU RUIII Installation Instructions Ver 3 21...
Page 130: ...PV9386 F4E005 18 4 In the following screen click the Yes button to complete the uninstallation...