4.3 MRC check setting
81
OER-Elite OPERATION MANUAL
Ch.4
When this setting is activated and the MRC check result is not input before start of the following
process, the reprocessor inhibits the start of the following process and message screen is displayed.
Reprocessing
Leaking Scope Decontamination
Water Line Disinfection
Self-disinfection & Water sampling
Detergent Line Disinfection
Alcohol Line Disinfection
NOTE
This setting is provided to prevent forgetting of the input of the MRC check result.
The MRC check result can be input even when this setting is disabled.
4.3
MRC check setting
1
Select “MRC Check Setting” on the settings menu.
2
Select setting from options.
No.
Item/Button
Setting options
Description
1
MRC Check Setting OFF
The reprocessing process starts even if the disinfectant solution
concentration check result is not input.
ON (default)
If the disinfectant solution concentration check result is not input
before start of the process, the reprocessor inhibits start of the
process and notifies that the check result is not input.
2
Cancel
–
Returns to the Setting menu without saving the selected setting.
3
Save
–
Returns to the Setting menu after saving the selected setting.
1
2
3