70
3.7 Checking the MRC level and entering the check result
OER-Elite OPERATION MANUAL
Ch.3
Required items
Table 3.2
NOTE
For the usable test strips, refer to Section 2.8, “Consumable accessories
(Optional)”.
Checking the MRC level
WARNING
Since the period of validity of the disinfectant solution varies depending on multiple
factors including the storage condition and the environmental temperature of the
reprocessor’s installation location, be sure to check the concentration of the
disinfectant solution before reprocessing endoscopes. If the concentration is below
the recommended minimum concentration or the validity date specified for the
disinfectant solution has expired, replace the disinfectant solution. Otherwise, the
endoscope reprocessing would become insufficient.
NOTE
If you collect the disinfectant solution from the drain port, the reduction of the
solution inside the disinfectant solution tank may make it impossible to start
reprocessing.
Check
Required Items
Chemical indicator (Test strip)
Clean cloth