4.16 Auxiliary water tube cleaning setting
OER-Elite OPERATION MANUAL
Ch.4
When this setting is activated, an auxiliary water tube (MAJ-855) can be reprocessed simultaneously
with an endoscope with water jet function by using dedicated connecting tube MAJ-2138. In this case,
two endoscopes with two auxiliary water tubes can be reprocessed in one reprocessing process.
To obtain optional connecting tube MAJ-2138, contact Olympus.
WARNING
If the setting is activated, always use connecting tube MAJ-2138 for an endoscope
with water jet function. Otherwise, channel connectivity monitoring may not be able
to detect improper connection of the connecting tubes.
NOTE
• If the setting is activated, the Connection guide screen which appears after reading
the scope ID displays MAJ-2138 as the connecting tube to be connected to the
water jet function.
• If the setting is activated, an endoscope with water jet function cannot be
reprocessed without an auxiliary water tube (MAJ-855).
• If this setting is disabled, either 2 endoscopes or 1 endoscope and 1 auxiliary water
tube (MAJ-855) can be simultaneously reprocessed.
• For details how to install endoscopes and auxiliary water tubes simultaneously,
refer to Section 6.6, “Loading of endoscopes and accessories”.
4.16 Auxiliary water tube cleaning setting
1
Select “Auxiliary Water Tube Cleaning Setting” on the Setting menu.
2
Select setting from options. Then press “OK” or “Save” button.