3.6 Entering ID
49
OER-Elite OPERATION MANUAL
Ch.3
NOTE
• When scope ID is entered using the software keyboard, the auto leak test and
channel blockage monitoring functions will not be available. In the manual leak test
function, the scope ID can be entered by the software keyboard. To enter the scope
ID, Press the “Edit” button during the manual leak test. Then, refer to step 2 in
“
Scope ID entry” on page 49.
• In the auto leak test function, the scope ID cannot be entered by the software
keyboard (on-screen). RFID of the endoscope is necessary to perform the auto
leak test.
• When User ID setting of Reprocessing (end of reprocessing) is activated, the user
ID can be entered by the software keyboard (on-screen) at the end of reprocessing.
To enter the user ID, press the “Edit” button at the end of reprocessing. Then, refer
to Step 2 in “
Physician ID, user ID, or procedure ID entry” on page 59. For details
of User ID setting, refer to Section 4.5, “User ID Setting”.
• When user ID setting of “Functions & Replacement of Consumable Items” is
activated, the user ID can be entered by the software keyboard (on-screen) before
starting each function or replacement. To enter the user ID, Press the “Edit” button
during each function or replacement. Then, refer to Step 2 in “
Physician ID, user
ID, or procedure ID entry” on page 59. For details of User ID setting, refer to
Section 4.5, “User ID Setting”.
Scope ID entry
1
Press the “Edit” button of scope ID on the ID
select screen.
Figure 3.17