910-269-EN, Rev. B, June 2011
296 Chapter 17
Figure 17-18 The first echo
8.
Choose
1/5 > PA Cal > Start
to begin the calibration process.
The screen freezes and the
Enter Value for Zero Cal
dialog box appears.
Figure 17-19 The Enter Value for Zero Cal dialog box
9.
Adjust the value to match the known thickness of the gated indication (4.000 in. in
this example), and then choose
Continue
to continue to the second calibration
step.
Downloaded from
Summary of Contents for EPOCH 1000 Series
Page 30: ...910 269 EN Rev B June 2011 18 Preface Downloaded from ManualsNet com search engine ...
Page 50: ...910 269 EN Rev B June 2011 38 Chapter 1 Downloaded from ManualsNet com search engine ...
Page 60: ...910 269 EN Rev B June 2011 48 Chapter 2 Downloaded from ManualsNet com search engine ...
Page 100: ...910 269 EN Rev B June 2011 88 Chapter 3 Downloaded from ManualsNet com search engine ...
Page 186: ...910 269 EN Rev B June 2011 174 Chapter 9 Downloaded from ManualsNet com search engine ...
Page 206: ...910 269 EN Rev B June 2011 194 Chapter 10 Downloaded from ManualsNet com search engine ...
Page 258: ...910 269 EN Rev B June 2011 246 Chapter 12 Downloaded from ManualsNet com search engine ...
Page 280: ...910 269 EN Rev B June 2011 268 Chapter 14 Downloaded from ManualsNet com search engine ...
Page 286: ...910 269 EN Rev B June 2011 274 Chapter 16 Downloaded from ManualsNet com search engine ...
Page 318: ...910 269 EN Rev B June 2011 306 Chapter 17 Downloaded from ManualsNet com search engine ...
Page 322: ...910 269 EN Rev B June 2011 310 Chapter 18 Downloaded from ManualsNet com search engine ...
Page 332: ...910 269 EN Rev B June 2011 320 Chapter 19 Downloaded from ManualsNet com search engine ...
Page 350: ...910 269 EN Rev B June 2011 338 List of Figures Downloaded from ManualsNet com search engine ...