910-269-EN, Rev. B, June 2011
214 Chapter 11
11.2.8
TVG Table
The TVG table option for the EPOCH 1000 Series is a powerful tool designed to
facilitate manual definition of TVG setups, including gain manipulation, fine gain
adjustment, and TVG point addition or deletion. The TVG table option only applies to
TVG curves. With the TVG table displayed, you cannot toggle between the DAC and
the TVG views. The TVG table function of the EPOCH 1000 Series visually displays
the slope of the change in gain across the defined screen range using a line. This
allows you to easily visualize the relationship between the indications and any
custom gain modifications made during the TVG table setup. The EPOCH 1000 Series
TVG table function has a number of primary uses discussed below.
Immersion Inspection
One of the main uses for the TVG table function is in immersion setups. A standard
TVG curve can be created using a reference block to aid in flaw sizing prior to
immersion inspections. However, the interface echo during an immersion inspection
can be very high in amplitude, which can severely decrease the near-surface
inspection capabilities of the instrument. Using a TVG table, you can manually adjust
the gain in the screen range right around the interface echo to suppress the effects of
the interface echo. Other echoes near the surface are then easier to see and size.
The EPOCH 1000 Series TVG table can be used at full PRF (6000 Hz) in
conjunction with the optional interface gate. This allows the EPOCH 1000 Series
to be used as a true immersion scanning instrument.
Manual TVG Setup
In many circumstances, a part must be inspected for flaws using a TVG curve to
determine the rejectability of any flaws found within the part, but the part is too large
or expensive to have a reference block of equal size and composition. The TVG table
feature allows you to manually add points along the TVG curve and manipulate the
gain slope across the screen range to create an accurate TVG curve without a reference
block (typically this is accomplished using the DGS diagram that corresponds to the
particular material and transducer being used).
Downloaded from
Summary of Contents for EPOCH 1000 Series
Page 30: ...910 269 EN Rev B June 2011 18 Preface Downloaded from ManualsNet com search engine ...
Page 50: ...910 269 EN Rev B June 2011 38 Chapter 1 Downloaded from ManualsNet com search engine ...
Page 60: ...910 269 EN Rev B June 2011 48 Chapter 2 Downloaded from ManualsNet com search engine ...
Page 100: ...910 269 EN Rev B June 2011 88 Chapter 3 Downloaded from ManualsNet com search engine ...
Page 186: ...910 269 EN Rev B June 2011 174 Chapter 9 Downloaded from ManualsNet com search engine ...
Page 206: ...910 269 EN Rev B June 2011 194 Chapter 10 Downloaded from ManualsNet com search engine ...
Page 258: ...910 269 EN Rev B June 2011 246 Chapter 12 Downloaded from ManualsNet com search engine ...
Page 280: ...910 269 EN Rev B June 2011 268 Chapter 14 Downloaded from ManualsNet com search engine ...
Page 286: ...910 269 EN Rev B June 2011 274 Chapter 16 Downloaded from ManualsNet com search engine ...
Page 318: ...910 269 EN Rev B June 2011 306 Chapter 17 Downloaded from ManualsNet com search engine ...
Page 322: ...910 269 EN Rev B June 2011 310 Chapter 18 Downloaded from ManualsNet com search engine ...
Page 332: ...910 269 EN Rev B June 2011 320 Chapter 19 Downloaded from ManualsNet com search engine ...
Page 350: ...910 269 EN Rev B June 2011 338 List of Figures Downloaded from ManualsNet com search engine ...