Specifications
Intended Use
The Olympic Bili-Lite
TM
Model 33/34 is a
floorstand phototherapy light intended to provide
therapeutic irradiance to treat neonatal
hyperbilirubinemia.
Model 34: The Olympic Bili-Timer
TM
is intended to
measure total phototherapy treatment time.
Irradiance
Average at 20 in. (51 cm)
Eight Olympic Blue (Type OB) bulbs:
11.9 ±4 µW/cm
2
/nm
Eight Super Blue (Type BB) bulbs:
22.4 ±7 µW/cm
2
/nm
Maximum at 20 in. (51 cm)
Eight Olympic Blue bulbs: 13.8 ±5 µW/cm
2
/nm
Eight Super Blue bulbs: 26.0 ±8 µW/cm
2
/nm
Dimensions
Size
64.8-in. high x 26-in. wide x 25-in. deep
165-cm high x 66-cm wide x 63.5-cm deep
Size (With Tilt)
66.3-in. high x 26-in. wide x 25-in. deep
168-cm high x 66-cm wide x 63.5-cm deep
Weight
Model 33:
67 lb
30 kg
Model 34:
69 lb
31 kg
Height Adjustment*
43.3–59.3 in. 110–151 cm
*Distance from bottom of light hood to floor.
Electrical
Power Requirements
Voltage: 120V~
Consumption: 3.0A
Frequency (current): 60 ±5Hz
Ground Impedance
Does not exceed 0.13
Ω
Fuses
3AB or Type ABC, 5A, normal blo, UL approved
or equivalent
Bulbs
Nominal bulb life (for 75% of initial irradiance):
3,000 hours
Environmental
Operating Temperature
50–85°F
10–29°C
Shipping Temperature
-40–158°F
-40–70°C
Storage Temperature
Model 33: -40–158°F
-40–70°C
Model 34: -4–113°F
-20–45°C
Relative Humidity (RH)
Operating: 10–95%, non-condensing
Shipping/storage: 10–100%, condensing
Pressure
0.5–1.0 atm
500–1600 hPa
Altitude (Maximum)
10,000 ft
3,000 m
Noise
Maximum noise level: 50 dB(A)