OLYMPIC BILI-LITE
Model 33/34 Instruction Manual
2-3
Assembling the Bili-Lite
4
Fully tighten the height adjustment knob, then cut the safety string.
5
Slowly and carefully release the height adjustment knob until the support post is
at its full extension, then tighten the knob again.
6
Remove the three knobs from the bolts on top of the light hood (see Figure 2.1).
7
Lift the light hood, then insert the bolts on top of the light hood through the
holes in the horizontal arm of the support post (see Figure 2.1).
8
Reinstall the knobs on the bolts and tighten.
9
Remove the plastic cord clamp from the support post (see Figure 2.1).
10
If you are adding a tilt assembly, perform the hood tilt assembly steps now (see
page 2-4). Once the hood tilt attachment is installed, continue to step 11.
11
Slip the power cord into the clamp and reinstall it on the support post. Leave
approximately 28 in. (71 cm) of slack in the cord above the clamp to allow
for adjustment.
12
Remove the safety shield by removing the retaining screws and washers. Then
gently slide the shield out (see Figure 2.3).
13
Remove the protective covering from the safety shield.
14
Install the bulbs: For each new bulb insert the prongs into the socket, then
carefully rotate the bulb into place.
15
Insert the safety shield, place the washers, then tighten the retaining screws.
16
Test for proper operation (see page 2-8).
WARNING
The spring inside the support post is compressed; releasing the height adjustment
knob will permit the support post to extend suddenly and forcefully to a distance of
16 in. (40.6 cm). Fully tighten the height adjustment knob before cutting the string.
While securely holding the support post, carefully release the knob.
CAUTION
The light hood is heavy and awkward. A second person should help support the light hood
while installing or removing it.
CAUTION
The safety shield must always be in place during phototherapy to protect the patient from
ultraviolet radiation and broken bulbs.