OLYMPIC BILI-LITE
Model 33/34 Instruction Manual
6-9
Repair Procedures
Figure 6.5
Bulb holder and cover
To install the new bulb holder:
1
Install the new bulb holder:
a
Place the new bulb holder into position.
b
Slide the two connectors under the connector screws, then tighten
the screws.
Figure 6.6
Resizing the connector for older models
c
Place the bulb holder cover over the bulb holder. Place a drop of Loctite 222
on the screw, then place the washer and tighten the nut to secure the cover
(see
2
Place wires 5 and 22 into the clip centered on the inside the hood.
3
Insert the reflector into the hood.
4
Place the washers, then tighten the nuts that secure the reflector to the
light hood.
5
Attach the fan shroud: Tighten the three Phillips-head screws and washers.
6
Attach the safety shield bracket: Tighten the two acorn nuts and two
Phillips-head screws and washers.
Ballast
Red connector
Connector screw
Bulb holder cover screw
Nut
Bulb holder cover
Black connector
Connector screw
Bulb holder
Starter
Starter socket
NOTE
#
The new bulb holder has a tighter area for the red connector. In older devices, to prevent
the bulb holder from cracking as the screw is tightened, it may be necessary to shorten
both tines of the red connector by 1/8 in. (0.3 cm)(see
Original length tine
Tine shortened 1/8 in.