78
Fault
Code
Fault Type
4-20mA
Output
Description
1FFF
Failed Sensor
1 mA
Smart sensor communication error
2FFF
Missing Sensor
1 mA
Sensor board communication error
ConF
Sensor
Configuration
1mA
Sensor internal parameters error –
Recover after factory configuration of
sensor
CFFF
Calibration
Failed
1 mA
Calibration
error
–
Recover
after
calibrating or replacing the sensor
0FFF
Zeroing Failed
1 mA
Zeroing error – Recover after zeroing or
calibrating
Table A - 2 Fault Code Description
Open Loop Condition
When any of the 4-20mA channels is
not being used it should be
terminated
by
inserting
the
specifically provided (250 ohms for
HART CH-1 and 100 ohms for
isolated CH-2) resistor between the
respective channel output terminal
and ground terminal.In case an
unused channel is not terminated
with the provided resistor, a ‘P’ will
appear at the status bit indicating
the open loop condition. Also in
case the channel output is being
used but one of the connecting
wires is damaged or disconnected,
same condition will be displayed to
let the user know about the
disconnection in the wiring. HART
communication
cannot
be
established
with
a
physical
disconnection under this condition.
Figure A - 13 Sensor 1 Open Loop
Condition Display
Summary of Contents for iTrans 2
Page 6: ...6...
Page 9: ...9...
Page 10: ...10...
Page 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Page 20: ...20...
Page 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Page 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Page 52: ...52...
Page 60: ...60...
Page 66: ...66...
Page 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Page 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Page 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Page 86: ...86...
Page 90: ...90...
Page 98: ...98...
Page 100: ...100...
Page 104: ...104...
Page 105: ...105...