
102
The propane IR sensor is factory configured to accurately monitor propane
gas. As seen in Figure F - 2 the cross-sensitivity of the propane IR sensor
does permit accurate measure of other combustible gases via a cross-
reference factor. It should be noted however, that the propane-IR sensor
WILL respond to other combustible gases and is not propane specific.
Figure F - 2 Cross-sensitivity chart for
Propane IR Sensor
The output of the IR sensor can be disrupted by sudden changes in
temperature. If there is an excessive change in the ambient temperature,
gas sample temperature or flow rate, then the output signal will be
monmentarily frozen. Correct operation is restored when the effects of the
transient have settled. Rates of change in the ambient temperature should
be restricted to 2°C/minute and gas flow rates kept below 0.6 L/minute.
Extreme pressure variations will cause errors in readings. The unit should
be recalibrated if the atmospheric pressure change is greater than 10% from
the orignial pressure.
# # #
Summary of Contents for iTrans 2
Page 6: ...6...
Page 9: ...9...
Page 10: ...10...
Page 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Page 20: ...20...
Page 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Page 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Page 52: ...52...
Page 60: ...60...
Page 66: ...66...
Page 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Page 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Page 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Page 86: ...86...
Page 90: ...90...
Page 98: ...98...
Page 100: ...100...
Page 104: ...104...
Page 105: ...105...