
38
As gas concentrations increase, the
respective channel’s readings will
respond accordingly. If low or high
alarm levels are exceeded, an alarm
indication will appear in the first
digit of the display. An “L” indicates
a low alarm while an “H” indicates a
high alarm. If a 4-20mA fault occurs,
either a “P” indicating an open loop,
or an “U” indicating 4-20 over-range
will be present. From the Normal
Operating Mode, the
can
enter into the program mode in one
of two ways.
Figure 5-3 Sample Low and High Alarm
Displays
To enter the Program Mode without opening the enclosure, pass over the
embedded reed switch located under CH1 with the magnetic wand (see
Figure 5-4). This will enter you into the non-intrusive program mode.
In this mode you can check sensor type, zero the unit, calibrate the unit,
change the span gas value, and view sensor span. With the enclosure top
removed, Program Mode can be entered using the “MODE” key. The
available functions are listed in Chapter 8 | Troubleshooting.
Summary of Contents for iTrans 2
Page 6: ...6...
Page 9: ...9...
Page 10: ...10...
Page 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Page 20: ...20...
Page 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Page 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Page 52: ...52...
Page 60: ...60...
Page 66: ...66...
Page 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Page 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Page 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Page 86: ...86...
Page 90: ...90...
Page 98: ...98...
Page 100: ...100...
Page 104: ...104...
Page 105: ...105...