50
NOTE:
Before the
will calibrate, the unit will enter the zeroing
process. Please make sure that you do not apply gas to the instrument while
it is zeroing.
The
will automatically zero before calibration. Zeroing is designated
with a flashing “0” in the status bit. Once zeroing is complete, the
will automatically enter the calibration routine. Calibration is designated
with a flashing “C” in the status bit.
After zeroing finishes, the
is ready to calibrate. When the flashing
“C” appears on the display, apply calibration gas. As the
responds
to the gas, the current reading will be displayed on the top line of the LED
display. To abort calibration at any time, press the “Mode” key.
NOTE:
Check and verify span setting before starting a calibration.
NOTE:
Please refer to Appendix D for a complete list of factory default span
gases.
NOTE:
Flow rate for calibration is 0.5 liter per minute (LPM) except for NH3,
ClO2, Cl2, NO2, SO2, and HCl which require 1.0 LPM.
Changing Span Gas Concentration
The span option is designated with a flashing “S” in the status bit with the
current span value next to it.
To change the span value, press the
“ ” key during the 10-second countdown. If you do not press the “ ” during
the 10-second countdown, the
will return to the Normal Operating
Mode.
Summary of Contents for iTrans 2
Page 6: ...6...
Page 9: ...9...
Page 10: ...10...
Page 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Page 20: ...20...
Page 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Page 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Page 52: ...52...
Page 60: ...60...
Page 66: ...66...
Page 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Page 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Page 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Page 86: ...86...
Page 90: ...90...
Page 98: ...98...
Page 100: ...100...
Page 104: ...104...
Page 105: ...105...