Figure 5-10 Apply CalGas Display
NOTE:
Flow rate for calibration is 0.5 liter per minute (LPM) except for NH3,
ClO2, Cl2, NO2, SO2, and HCl which require 1.0 LPM.
Changing Span Gas Concentration
The option after calibration is Span
Gas Concentration. The span option
is designated with a flashing “S” in
the status bit with the current span
value next to it. To change the span
value, place the magnetic wand over
CH2
during
the
10-second
countdown. If you do not place the
magnet over CH1 during the 10-
second countdown, the
will
return to the Normal Operating
Mode. If you initiate the change span
option, the status bit will start to
flash and the
span value can
now be changed.
Figure 5-11 Sample Span Gas
Concentration Display
The current span value is displayed
on the top line of the LED display.
To increment the span value, pass
the magnetic wand over CH1. When
the desired value is reached, pass
the magnetic wand over CH2 to
accept and save changes. Passing
over CH1 or letting the timer count
down to zero without saving the new
value, will take you back into the
Programming Mode.
Figure 5-12 Flashing Status Bit
NOTE:
Span Gas Concentration for combustibles can be set from 0% to
100%LEL. For the sake of resolution, the Span Gas Concentration should be
set above 20% LEL.
Summary of Contents for iTrans 2
Page 6: ...6...
Page 9: ...9...
Page 10: ...10...
Page 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Page 20: ...20...
Page 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Page 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Page 52: ...52...
Page 60: ...60...
Page 66: ...66...
Page 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Page 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Page 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Page 86: ...86...
Page 90: ...90...
Page 98: ...98...
Page 100: ...100...
Page 104: ...104...
Page 105: ...105...