55
Addr
Inst
R/W
Host
R/W
Range
Description
Examples:
+5 ppm = register value of 00005
10
= $0005
-5 ppm = register value of 65531
10
= $FFFB
40103
R*
R*
MSB = $01 to
$FF
LSB = $01 to
$FF
Gas Type
Holds the decimal place holder and the
gas type code. The most significant byte
(MSB) holds the number of decimal places
to be used in calculations for this gas.
This decimal locator applies to all
subsequent values of gas readings within
other registers. This can be read by the
instrument. The least significant byte
(LSB) holds a code which identifies the
gas type. This can be read by the host.
MSB = Decimal place holder $01 to $FF
LSB = Gas type code $01 to $FF
$01
CO
Carbon Monoxide
$02
H
2
S
Hydrogen Sulfide
$03
SO
2
Sulfur Dioxide
$04
NO
2
Nitrogen Dioxide
$05
Cl
2
Chlorine
$06
ClO
2
Chlorine Dioxide
$07
HCN
Hydrogen Cyanide
$08
PH
3
Phosphine
$09
H
2
Hydrogen
$0B
CO
2
Carbon Dioxide
$0C
NO
Nitric Oxide
$0D
NH
3
Ammonia
$0E
HCl
Hydrogen Chloride
$14
O
2
Oxygen
$15
CH
4
Methane
$16
LEL
Lower Explosive Limit
(Combustible Gases)
$17
C
6
H
14
Hexane
$1A
C
5
H
12
Pentane
$1B
C
3
H
8
Propane
$4D
C
2
H
6
O
Ethanol
$50
C
2
H
4
Ethylene
$6F
C
3
H
6
Propylene
$C9
C
4
H
10
Butane
Examples:
Summary of Contents for iTrans 2
Page 6: ...6...
Page 9: ...9...
Page 10: ...10...
Page 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Page 20: ...20...
Page 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Page 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Page 52: ...52...
Page 60: ...60...
Page 66: ...66...
Page 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Page 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Page 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Page 86: ...86...
Page 90: ...90...
Page 98: ...98...
Page 100: ...100...
Page 104: ...104...
Page 105: ...105...