32
Set the eddy current probe on the sample with Cond 2 conductivity and
press
.
Figure 21 –
Entering the conductivity value for the sample Cond 2
To put on the sample with Cond 1 conductivity the gap element (20 mil
[0,5 mm]), set the eddy current probe on the sample and press
.
Figure 22 –
Entering the gap value for setting on the Cond 1 sample
Measuring Conductivity and Coating Thickness with the Eddycon C
Guidelines for successful operation:
For accurate measurement of conductivity the surface coating
thickness should be 0.5 mm (0.020 inches).
The surface to be measured should be flat, or of the same curvature as
the calibration standards. Where curved surfaces must be measured,
additional error may be introduced.