27
"
:"
parameter variation allows change turning angle of threshold level
relative to the complex plane center. Turn angle variation is carried out in the
range from 0˚ to 359˚ with a step of 1˚, 10˚, 100˚ (see Figure 16).
"
:"
parameter variation allows change turning angle of threshold level.
Turn angle variation is carried out in the range from 0˚ to 180˚ with a step of
1˚, 10˚, 100˚(see Figure 16).
Select the frame polarity "Pol:", which makes the alarm trigger:
– the signal gets beyond the limits of the threshold level;
– the signal is within the limits of the threshold level.
If the symmetrical displaying of threshold level on the position “Symm:”
with
,
keys is required, this parameter can be activated.
Setup of "Trapezium" type frame
To set up trapezium type frame, operator should select "Type: " in
special item of settings ALARM–
(see Figure 17).
Figure 17 –
Threshold level setup "Type: "
"X1:"
parameter variation allows changing the angle position of
threshold level in X1-direction. X1 parameter variation is carried out in the
range from - 32768 to 0 with a step of 1, 10, 100, 1000, 10000 (see Figure 17).
"X2:"
parameter variation allows changing the angle position of
threshold level in X2-direction. X2 parameter variation is carried out in the
range from 0 to 32768 with a step of 1, 10, 100, 1000, 10000 (see Figure 17).
"Y1:"
parameter variation allows changing the angle position of
threshold level in Y1-direction. Y1 parameter variation is carried out in the
range from - 32768 to 32768 with a step of 1, 10, 100, 1000, 10000 (see
Figure 17).
Y2
Х2
Х1
Y1