11
Figure 2. H101-CRYO-BL Standard Connection
1.
Connect the Water Bath CRYO to the
port labeled ‘
Water Bath
’ on
the rear panel of the T-Unit (see
2.
Plug the NTC Connector
on the port labeled ‘Ambient Temperature’ located on the rear panel of
the
T-Unit (see Figure 2). The NTC thermistor can be also used to monitor the room temperature when
working in Sample Feedback Mode.
3.
The Fine Gauge Thermocouple (green) plugs into the green port labeled
‘Reference Temperature’
located on the rear panel of the T-Unit (see Figure 2). W
hen operating the device in ‘Sample
Feedback
Mode’ place
this thermocouple in a Reference Well filled with water. Make sure that the
end of the Fine Gauge Thermocouple is fully immersed.
Tip
► The
Fine Gauge Thermocouple may be placed into a Reference Well
if running the “Sample
Feedback
Mode
” (see paragraph
6.2.1.1). The Fine Gauge Thermocouple may be left in air if the system is
operated in ‘
Chamber feedback
mode’
(see paragraph 6.2.1.1).
4.
Connect the touch screen interface OKO-TOUCH to the T-Unit using the serial cable provided (see