25
6.2.2.5
Alarms
This page allows you to set deviation values (from set-point) beyond which you want the system to trigger
an Alarm.
Figure 30. How to enter in Alarms page
Tip
►
Flag
“Buzzer” if you want the Alarm to be acoustical as well rather than just being displayed.
Figure 31. How to set the Deviation and Time Alarm
Figure 31. In this example the following settings are displayed (valid only after the system has reached its
stationary state): if Temperature Deviation from set point is 1°C or greater (i.e. if set point temperature is
37°C and the temperature reaches a value equal or less than 36.0°C or equal or more than 38.0°C) for a
period of time equal or longer than 15 minutes (“Temp time“ set in this exampl
e) then the system triggers
an Alarm.
6.2.2.6
Logging
To use this system feature, refer to the Data Logging quick instructions.
Figure 32 shows an overview of the complete system displaying current values of all controlled parameters.
Similarly to the Status Page you may have to refer to this window when asking for Okolab technical support.