6.3 Select Measurement Resolution
If the sensor unit is located outside of its start position at the upper end of the stand, it first moves
back to its initial position (the LEDs light up red).
Select the resolution for the intended measurement in the window (figure 18):
Following resolutions are available:
Low resolution
Quick measurement,
Measurement duration: approx. 8 minutes
Medium resolution
Measurement duration: approx. 11 minutes
High resolution
Very detailed measurement,
Measurement duration: approx. 14 minutes
Select one of the available resolutions and confirm with
OK
!
6.4 Perform Calibration / Zero Adjustment
Before the actual measurement takes place, a so-called zero adjustment is performed. During this
process the current magnetic field of the surrounding environment is measured. Due to these
reference values, the subsequent body measurement can be visualized correctly.
As soon as the resolution (see section 6.3 “Select Measurement Resolution“ on page 25) or the
magnetic field of the environment has been changed (e.g. by adding, moving or removing room
inventory), a new zero adjustment must be processed.
If the last zero adjustment took place 2 or more hours ago, we recommend a new zero adjustment.
This is indicated in a message (shown in figure 19).
24
Figure 18: Select measurement resolution
Figure 19: Recommendation for executing zero adjustment