![Ohio Medical Corporation MiniOX 1 Operating Manual Download Page 13](http://html1.mh-extra.com/html/ohio-medical-corporation/miniox-1/miniox-1_operating-manual_3515043013.webp)
P/N 711462 (Rev.3) 02/2016
Recommended separation distances between portable and mobile RF communications equipment
and the MiniOX
®
1
The MiniO
X
®
1
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the MiniO
X
®
1
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the MiniO
X
®
1
as recommend-
ed below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d
= 8.8
S
P
80 MHz to 800 MHz
d
= 1.2
S
P
800 MHz to 2.5 GHz
d
= 2.3
S
P
0.01
0.88
0.12
0.23
0.1
2.8
0.38
0.73
1
8.8
1.2
2.3
10
28
3.8
7.3
100
88
12
23
MiniO
X
®
1
Oxygen Analyzer
Section 3, Specifications
3-4