Electromagnetic Compatibility (EMC)
Ensure that the specified electromagnetic environment and separation distances for the MiniOX
®
1 Oxygen
Analyzer observed according to the tables below. Misuse may result in failure to perform as designed,
including incorrect/erratic readings.
Guidance and manufacturer’s declaration - electromagnetic emissions
The MiniO
X
®
1
is intended for use in the electromagnetic environment specified below. The customer or the user of the
MiniO
X
®
1
should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The MiniO
X
®
1
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Group B
The MiniO
X
®
1
is suitable for use in all establishments
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Not applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
Not applicable
Guidance and manufacturer’s declaration - electromagnetic immunity
The MiniO
X
®
1
is intended for use in the electromagnetic environment specified below. The customer or the user of the
MiniO
X
®
1
should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+6kV contact
+8kV air
+6kV contact
+8kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast transient/burst
IEC 61000-4-4
+2kV for power supply lines
+1kV for input/output lines
Not Applicable
Not Applicable
Surge
IEC 61000-4-5
+1kV line(s) to line(s)
+2kV line(s) to earth
Not Applicable
Not Applicable
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Not Applicable
Not Applicable
Not Applicable
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
MiniO
X
®
1
Oxygen Analyzer
Section 3, Specifications
WARNING
3-2