Troubleshooting Screen
The
TROUBLESHOOTING
screen was designed to test each set of sieve beds
individually to see if one is out of purity in the system. It is important to note that once
you enter this troubleshooting mode the system will not continue to operate normally.
Once the troubleshooting has been completed the system can be re-started again.
To begin the troubleshooting you need to push the button under the sieve bed set
you want to start up and then push the
ON/OFF
button. This will only activate that set of
sieve beds. Then, after 15 minutes of running, you will see on the „
Oxygen Purity
Gauge
‟ if
that sieve bed set is not running properly (making at least 90% pure oxygen).
In some cases, especially if the reading is only slightly low (80-89%) a leak may
have developed in the machine. While the system is still running and pressurized, a small
amount of soapy water can be used to try to find the leaking part. If the purity is very
low the sieve beds may need to be replaced. They are designed to be able to be replaced
quickly and easily in the field. If you do not have a replacement set of beds in your
stock, contact the factory or your local distributor.