Effective 11/2010
38
Graph Screen
This screen shows the graph based on the readings on the
GAUGE
screen.
60
minutes are
displayed at a time and 24 hours are stored in memory. If power to the machine is lost
(
Emergency-Stop
button is pushed in, machine is unplugged, etc.), the chart memory will be
lost. The
GRAPH
screen shows the history of what has been displayed on the
GAUGE
screen for up to the previous 24 hour period. Note that the machine only has a 24 hour
memory when it is
ON
so after 24 hours it will start deleting or writing over previous history.
In addition, these readings are stored in dynamic memory, so if the machine is de-energized,
these readings will be lost. The readings that are recorded are for: Oxygen Purity, Bed
Oxygen Pressure, Air Tank Pressure and Oxygen Tank Pressure. This information can be
stored indefinitely when downloaded to computer with an optional touch screen SD card.