48
Diaphragm
(#120-58-23)
Space for
Thermocouple
Retaining Ring
(#120-58-38)
9. Apply high temperature grease to the threads, seal ring, o-ring, and the
retainer ring.
5. Insert the diaphragm into the cell cap so the gap clears the thermocouple.
Insert the other retaining ring into the groove above the diaphragm so the
gap clears the thermocouple as shown below.
6. Place the metal seal ring on the external portion of the cell
cap with the narrow end pointed toward the threads. The
seal ring should not need to be removed from the top cell
cap. If it must be replaced, the metal seal ring must be
reinstalled in the same orientation.
7. Place an o-ring on the cell cap above the metal seal ring.
8. Install the retaining ring above the o-ring.
Metal Seal Ring
(Cross Section)
Thermocouple