5. Plug the thermocouple from the cell cap to the port on the cabinet leaving
the thermocouple gland slightly loose on the cell cap.
6. Turn the "WATER ON/FILL" switch to the on position until water comes
out of the thermocouple gland and ventilation port on the top cell cap.
7. Once the water comes out of the ventilation port, tighten the
thermocouple gland completely and turn the Water On/Fill switch to the
"OFF" position. This will ensure that all air has been purged from the cell.
8. Clean and dry the top cell cap and cabinet of excess water.
9. Make sure all the cables are connected properly. See page 11 for more
information.
⅝" Glands
Water Supply
Tube
Transducer Cable
Spring Holder
Test Cell
Thermocouple
5/32
" Allen Head
⅝" Gland