OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
75
6. Once the reading on the dry block stabilizes, click accept.
7. Now set the dry block to 200°F.
8. Enter “Setpoint 2” at 200°F (93.3°C) and wait until the reading reaches
the target temperature.
9. Once the reading on the dry block stabilizes click accept.
10. Now set the dry block to 400°F.
11. Enter “Setpoint 3” at 400°F (204.4°C) and wait until the reading reaches
the target temperature.
12. Once the reading on the dry block stabilizes click accept.