OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
74
Calibration requires the use of a calibrated heating device. For further
assistance, contact OFITE Technical Support.
1. Start the UCA or SGSM software.
2. Connect the thermocouple to a dry block calibrating device.
3. Set the dry block to 100°F (37.8°C).
4. Access the Thermocouple
Calibration: Utilities
→
Calibrate
→
Thermocouple.
A warning prompt will appear to
inform the user that all parameters
will be erased if the entire calibration
process is not completed.
5. Enter “Setpoint 1” at 100°F (37.8°C) and wait until the reading reaches
the target temperature.
Calibration
Thermocouple