OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
65
Symptom
Cause
Remedy
Software
There is no
transit time
The transducers are dirty
Too much couplant has been
applied to the transducers
The transducer cables have
been damaged
The transducers have
exceeded their useful life
Clean the transducers
Wipe off the transducers
and apply a smaller
amount of couplant to the
transmitter surface
Replace the transducer
cables
Replace the transducers
Trace lines are
missing from the
graph
The boxes for each line are
unchecked
Check the appropriate
boxes
Leaks
Water is not
flowing to the
cell
The filter is plugged
The fill tubing is plugged
The fittings and/or cell cap
openings are clogged with
cement
Clean the filter (page
Check the tubing for
obstruction
Check the openings and
clean if necessary
Water is leaking
from the fitting
The threads have been
damaged
The collar on the
thermocouple connection is
too high or too low
Cement has clogged the
inner surface of the port
Replace the fitting
Adjust the collar until 2
threads are exposed
Clean cement out of the
port