Quick Start
UCA
1. Set up the test in the software (page 12) or in the onboard display (page
24).
2.
Create a Test Profile in the software (page 18) or in the onboard display
(page 25).
3. Mix the cement slurry.
4.
Assemble the test cell (page 30), fill the cell with cement, load it into the
heating jacket, and make all the necessary connections (page 33).
5. Turn the Water Supply valve to “FILL” and wait for water to come out of
the thermocouple port.
6. When the cell is full, turn the tighten the thermocouple and turn the Water
Supply valve to “OPERATE”.
7. Set the “HEAT”, “COOL”, and “PRESSURE” switches to “AUTO”.
8. Click or touch the “Start Test” button.
9. After the test, allow test cell to cool down and remove it from the heating
10. Disassemble the test cell and remove the hardened cement (page 39).
11. Clean and inspect all of the test cell components.