15. Install the SGSM base to the mounting adapter as shown below. Screw
the threads facing inward and the allen heads on the outside of the base.
16. Insert the magnet shaft assembly (#120-58-010) into the SGSM base,
threads first with the magnet facing upward.
17. Lightly grease the threads of the mounting adapter that secure the
magnet cover.
18. Place the magnet cover over the magnet and screw it tight.
19. Lightly grease the threads that secure the plug and gland.
20. Install the plug and gland. Tighten them, then loosen them ¼ of a turn.
21. Secure the top cell cap in a vise with the threads facing down.
22. Carefully insert the assembled base and bob shaft through the top of the
cap, with the threads of the bob shaft going in first.
23. Once the base reaches the top of the cell cap, screw the assembly onto
the cell cap.
SGSM Base
Low Pressure SGSM
Thermocouple
(#120-51-020)
Vane Paddle
Assembly
(#120-53-01)
Magnet Shaft
Assembly
(#120-53-010)
Top Cell Cap
(#120-53-03)
Magnet Cover
(#120-53-09)
Plug
(#120-50-058)
Gland
(#122-004-002)
Magnet Cover
(#120-53-09)
SGSM Base
(#120-53-13)
Weep Hole