Page 2
PERIODIC MAINTENANCE PROCEDURES
COMMENTS
TECHNIQUE DATA
Circle One: 60 Hz or 50 Hz
Record X-ray Calibration Data
12-040-9/96
Periodic Maintenance
MINI6600 IMAGING SYSTEM
Perform System Diagnostic Tests
Verify the System Controls
Perform Accessory Operational Test
Inspect/Adjust Gen-Lock Calibration
Perform Camera and Beam Alignment
Record Image Resolution
Check Entrance Exposure
Actual Dose Rate = _______
FIELD SIZE 60 Hz MIN lp/mm
MEASURED
RECORD
NEW IMAGE
RESOLUTION
4"
6"
2.5 mm
3.0 mm
50 Hz MIN lp/mm
2.2 mm
COMMENTS:
Attach Films
PERIODIC MAINTENANCE PROCEDURES
COMMENTS
TP14 (GND) TP25
TP15
TP25
TP25
TP15
75 kVp @ 98.0 micro amp
58 kVp @ 35.2 micro amp
58 kVp @ 35.2 micro amp
40 kVp @ 19.6 micro amp
75 kVp @ 98.0 micro amp
TP14 (GND)
TP14 (GND)
TP14 (GND)
TP14 (GND)
+7.20 VDC to +7.80 VDC
+4.70 VDC to +5.10 VDC
+5.57 VDC to +6.03 VDC
+1.69 VDC to +1.83 VDC
+3.84 VDC to +4.16 VDC
VDC
VDC
VDC
VDC
VDC
TP14 (GND) TP15
40 kVp @ 19.6 micro amp +0.94 VDC to + 1.02 VDC
DOSIMETER
72 kVp to 78.0 kVp (+/- 4%)
75 kVp @ 98.0 micro amp
VDC
kVp
5. ANNUAL PM
Indicated vs. Actual Technique Verification
6. CLEAN-UP
7. COMPLETE PM FORM
END SEMI-ANNUAL PM
Accessing 6600 Controller PCB Test Points
2.6 mm
TEST POINTS
TOLERANCE
MEASUREMENTS
X-RAY VERIFICATION TEST DATA
( 0.89 R/min)
£