
MINI6600
TM
Mobile Digital C-Arm Service Manual - Periodic Maintenance
15. Verify the AVERAGING operation.
No Lamps Lit
Intrinsic Averaging
2 Frames
1 Lamp Lit
Low Averaging
4 Frames
2 Lamps Lit
Medium Averaging
8 Frames
3 Lamps Lit
High Averaging
16 Frames
TABLE 5
16. Verify the PATIENT ANNOTATION operation.
17. Verify the DISK VIEW operation (16 image option systems).
18. Verify the TEST PATTERN operation.
19. Verify the ALARM RESET operation.
20. Verify L-R operation.
21. Verify the HELP operation.
22. Verify the SET-UP OPTIONS operation.
NOTE:
The image directory, markers, X-ray summary, special apps and customize features are not
operational. Attempts at selecting these features will produce a negative beep.