Power Up the System
1. Plug the AC power cord into the wall receptacle.
2. Turn or press the System keyswitch/pushbutton to ON.
3. The system will perform a Power ON self-test during power up.
Cart Progress and Error Codes
1. Failure of the self-test will be indicated by error codes on the System rear panel LED’s.
2. If any code remains and the system fails to complete the Power ON sequence, refer to your Service
Manual (00-878315).
3. Verify that the elapsed time indicator is running.
NOTE:
The elapsed time indicator will increment every 6 minutes or 1/10th of an hour. If the indicator is
not running, initiate repairs before continuing Periodic Maintenance.
System Virus Inspection
1. Turn or press the system keyswitch/pushbutton to the OFF position.
2. Access DOS by inserting a DOS disk in drive A and turn or press the keyswitch/pushbutton to the ON
position.
3. When you see A:\> remove the DOS Disk and insert the Virus Scan Disk.
4. Run the virus scan program.
5. Repair/clean any files found infected. Remove the virus scan disk.
6. Reboot the system.