MINI6600
TM
Mobile Digital C-Arm Service Manual - Periodic Maintenance
Annual Maintenance
WARNING:
Steps within this procedure produce X-rays. Use appropriate precautions.
Complete all semi-annual PM procedures.
section of the Service Manual (00-878315). Complete the following sections:
Accessing 6600 Controller PCB Test Points
Indicated vs. Actual Technique Verification
Record Technique Verification readings on the PM Form.