MINI6600
TM
Mobile Digital C-Arm Service Manual - Periodic Maintenance
Inspect the Service Outlet Strip
1. Check for 115 VAC
±
5 VAC at the service outlet strip located below the monitors at the rear of the
System.
2. Check for 230 VAC
±
10 VAC at the service outlet strip located below the monitors at the rear of the
System.
Perform the System Diagnostic Tests
1. Insert the Service Disk into the disk drive. Same service disk as used for the Series 9600 C-Arm (00-
877444).
2. Press SETUP OPTIONS and access Level 2 diagnostics. Select the Setup Options menu, and then
select Run System Diagnostics. The following tests will appear:
•••••
Front Panel Test
•
System Memory Tests (AT Motherboard Memory)
•
Image Processor Tests (IP PCB Memory)
•
Toggle Motion Artifact (Toggles ON/OFF), Non Functional
3. Select the Front
Panel Test and press the following switches and verify that a hex code is displayed
and an audible tone is heard for each switch pressed: Negate, VCR Play, VCR Record, Auto, Help,
Brightness and Contrast, Fluoro, One Shot, Enhanced. Averaging, L-R.
4. Verify that the remaining tests pass successfully. Press the ENTER key to toggle through tests.
5. Remove the Service Disk from the disk drive.
6. Pressing ESC will exit the System Diagnostic tests and reboot the System.